Pipette Service , Calibration & Repair


Abimed is a leading provider of pipette service and maintenance. The latest state of the art balances are used both within our purpose built laboratory and in the customer workplace Abimed's experience and abilty to react to changing customer needs ensure unrivalled service and data reporting plus great value for money Some of the Abimed benefits
  • Service levels to match customer requirements
  • All major manufacturers pipettes-single, multichannel and electronic
  • Better reporting of results
  • Service at Abimed service centre, customer workplace or a combination of both
  • No limit on numbers, 5 day turnaround
  • ISO8655 compliant

For quality insurance for your measurement results and products your pipettes should be serviced and calibrated on a regular basis. In the Abimed pipette centre we offer a wide range of service levels for all manufacturer's single - and multichannel pipettes.

Service-levels for single channel pipettes variable and fix volume

Standard (S)Maintenance/Repair*
Cleaning internal and external parts, repair if necessary calibration 2 x 4 readings, Quick-Check, documentation Euro 25,90

Certificate (Z)Maintenance/Repair*
Cleaning internal and external parts, repair if necessary, calibration 3 x 10 readings according to EN ISO 8655 documentation Euro 38,80

Readings as received + Certificate (EZ/AZ)Maintenance/Repair*
Readings as received + Certificate (EZ/AZ) 3 x 10 readings as received, cleaning internal and external parts, repair if necessary, calibration 3 x 10 readings according to EN ISO 8655, documentation Euro 68,50

Service-levels for multichannel pipettes.

Standard (M-S) List Price
Cleaning internal and external parts, repair if necessary
Calibration 2 x 4 readings each channel, Quick-Check
Euro 25,90
Euro 2,45/Channel

Certificate (M-Z) List Price

Cleaning internal and external parts, repair if necessary
Calibration 3 x 10 readings each channel accordingto EN ISO 8655, documentation

Euro 38,80
Euro 5,40/Channel

* Spare parts for single- and multichannel pipettes will be charged separately.

Calibration sticker is fixed to each pipette detailing date of next service.

On request our sales people pick up your pipettes for repair personally.

The order form for pipette service can be downloaded to be attached to your pipettes.

Liquid Handling
Pipette Service
Chromatography
 
Cleaning Steps:
Cleaning steps
  • Ultrasonic bath
  • Rinsing with clear water
  • Chemical decontamination
  • Rinsing with clear water
  • Deionized water bath
  • Drying in special incubator
  • Cooling to ambient temperature

Service, maintenance & repair
Service, maintenance & repair
  • Disassembly of pipette
  • Examination of internal and external parts
  • Repair if necessary
  • Replacement of missing, broken or worn parts

Calibration and Documentation
Calibration and Documentation
  • Gravimetric evaluation of pipette performance
  • Adjustement where required
  • Calibration limits in accordance with manufacturer's specifications
  • Calibration sticker fixed to each pipette detailing date of next service
  • Service, documentation report and record of all measurements against serialised unit